Optics is our core value

  +603-76136915   P1-30 Lumi Tropicana, Persiaran Tropicana, 47800

Evaluation environment
Coaxial tube:  Kohler illumination type    /   Critical illumination type 
Objective lens: PEIR Plan 10X
Camera:BU406M(2048 x 2048, 5um)
Light source: SIS-150-AIR + band pass 1100nm
Inspecting object: Mirror

<Definition of the uniformity>
The average brightness in red square is defined as criteria of 100% uniformity.

 

<Original image /  Kohler illumination type>                                                                                             

 

 

Pixels (2048 ,2048)
Mean of uniformity of 8 blue squares(%)  = 82.20578420281981
Max uniformity of 8 blue squares(%) = 90.36996901226476
Min uniformity of 8 blue squares(%) = 76.87275048736467
Standard deviation= 5.101972294868123

 

 

 

 

<Original image /  Critical illumination type>   

       

Pixels (2048 ,2048)
Mean of uniformity of 8 blue squares(%)  =81.95705301193571
Max uniformity of 8 blue squares(%) = 95.12371239487848
Min uniformity of 8 blue squares(%) =68.47193763278311
Standard deviation=10.292410912360923

 

 

 

 

 

<Gaussian distribution>

This graph shows that the uniformity of Kohler illumination type is so stable comparing Critical illumination type.

 

 

Kohler illumination type

Critical illumination type

The joy of being useful to the world and the real thrill of manufacturing. That is the charm of Seiwa.

Copyright © 2024 Seiwa Optical Malaysia Sdn Bhd

This is a staging environment